Advertisement

NÜVE SANAYİ MALZEMELERİ İMALAT VE TİCARET A.Ş.
NB 5 / NB 9 / NB 20
WATER BATHS
USER'S MANUAL
Z14.K 25 255
Rev. No: 11
Rev. Date: 07/2017

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the NB 5 and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for NUVE NB 5

  • Page 1 NÜVE SANAYİ MALZEMELERİ İMALAT VE TİCARET A.Ş. NB 5 / NB 9 / NB 20 WATER BATHS USER’S MANUAL Z14.K 25 255 Rev. No: 11 Rev. Date: 07/2017...
  • Page 2 Lifetime of the device is 10 years. NÜVE SANAYİ MALZEMELERİ İMALAT VE TİCARET A.Ş. Saracalar Mah. Saracalar Kümeevleri No: 4/2 Akyurt 06750 Ankara / TURKEY : (90.312) 399 28 30 (pbx) : (90.312) 399 21 97 Sales : sales@nuve.com.tr Technical Service: nuveservice@nuve.com.tr...
  • Page 3 INFORMATION CONTAINED IN THIS DOCUMENT IS THE PROPERTY OF NÜVE. IT MAY NOT BE DUPLICATED OR DISTRIBUTED WITHOUT PERMISSION. PLEASE REGISTER ONLINE TO VALIDATE YOUR WARRANTY: To register your warranty online, please visit our webpage www.nuve.com.tr and fill in the “Warranty Registration Form”.
  • Page 4: Table Of Contents

    PERIODIC MAINTENANCE ……………………………………………………………………………………………………… 18 CLEANING …………………………………………………………………………………………………………………………….. 18 DISPOSAL MANAGEMENT CONCEPT ……………………………………………………………………………………………. 18 TROUBLESHOOTING …………………………………………………………………………………………………………………… 18 ERROR CODES EXPLANATIONS ……………………………………………………………………………………………… 19 FUSE REPLACEMENT ……………………………………………………………………………………………………………… 20 ELECTRICAL CIRCUIT DIAGRAM …………………………………………………………………………………………………. 21 10.1 NB 5 / NB 9 / NB 20 ELECTRICAL CIRCUIT DIAGRAM …………………………………………………………. 21...
  • Page 5: Introduction

    1. INTRODUCTION 1.1 USE AND FUNCTION NB 5, NB 9 and NB 20 Water Baths accurately control the temperatures of samples and commonly used in microbiology, research and industrial laboratories. They offer excellent temperature control of liquid producing a uniform and stable heating environment for the applications. They provide temperatures between 5°C above the ambient temperature and 99.9°C for procedures...
  • Page 6: Technical Specifications

    1 line. 2.2 OPTIONAL ACCESSORIES K 04 290 Plexiglass lid for NB 5 (resists up to 60°C) K 52 010 Stainless steel lid for NB 5 K 04 286 Plexiglass lid for NB 9 (resists up to 60°C) K 52 007 Stainless steel lid for NB 9 K 04 289 Plexiglass lid for NB 20 (resists up to 60°C)
  • Page 7: Precautions And Limitations On Use

    3. PRECAUTIONS AND LIMITATIONS ON USE The user shall pay attention to the following:  Do not operate the instrument for purposes other than its main purpose.  The instrument should only be used by authorized and trained staff after the instruction manual has been read carefully.
  • Page 8: Installation

    NB 5 Fuse (2x4A) NB 9 Fuse (2x5A) NB 20 Fuse Earthed Wall (2x10A) Sockets 5. INSTALLATION 5.1 ENVIRONMENTAL CONDITIONS The instrument is designed to operate safely under the following conditions:  Indoor use only  Ambient temperature: 5°C to 40°C ...
  • Page 9: General Presentation

     Check that the positioning of the device prevents interference with other equipment in the near surrounding. 5.6 GENERAL PRESENTATION DISPLAY AND CONTROL PANEL INPUT USB Figure 1 TANK TEMPERATURE SENSOR GRID Figure 2 ON/OFF SWİTCH WATER DISCHARGE SUPPLY INPUT AND LINE FUSES Figure 3...
  • Page 10: Control Panel

    5.7 CONTROL PANEL Figure 4 6. “Operating the program” lamp 1. Temperature display 11. Back / Alarm mute button 2. Time display 7. Data transfer lamp 3. Encoder button 8. Alarm led 9. “End of the program” lamp 4. Start / Stop button 5.
  • Page 11: Prior To Operation

    03-Encoder Button The encoder button has two physical movements. The button turn clockwise and anticlockwise to increase or decrease the temperature and time values of the program. Also, the button press for select / confirm. 04-Start / Stop Button Used button to start the device to operating at set values or to stop the operation. 05-Heating Lamp The led is “on”...
  • Page 12: Filling The Tank

    5.8.1 FILLING THE TANK MAX. Line Fill the tank with distilled water or liquid up to the maximum line MAX (see Figure 5). Undesired problems may occur if you fill up to the level lower than the max level. Figure 5 Make sure that the filled liquid is not flammable or explosive at the operation temperature.
  • Page 13: Op Operator Menu Parameters

    Wait by pressing the encoder button. Lift your hand when you see "oP" on the temperature display and again press the encoder button. If the device has a menu protection password, password screen “oPS” will come on the temperature display. You enter the set password to turn the encoder button right and left.
  • Page 14: Programming Summary

    7: Date Setting - Month: The month information of date is displayed. If updating is necessary, change. 8: Date Setting - Day: The day information of date is displayed. If updating is necessary, change. 9: Time Setting - Hour: The hour information of time is displayed. If updating is necessary, change.
  • Page 15 See the parameter flashing on the temperature display. By turning the encoder button set operating temperature value. Push the encoder button and save set value. See the parameter flashing on the temperature display. By turning the encoder button set operating Set alarm value.
  • Page 16 See that second LED flashes in the time display, again push the encoder button. See the parameter flashing on the time display. By turning the encoder button set operating time value (01 minute to 99 hours 54 minutes or Hold). Push the encoder button and save set value.
  • Page 17: Completion Of Operation

    In order to display the set values during the operation, push the encoder button once. The values set on the temperature display and the time display of the device will appear for 5 seconds.  During the operation of the program, the time starts to count up after the instrument has reached to the set temperature 6.3 COMPLETION OF OPERATION ...
  • Page 18: Periodic Maintenance And Cleaning

    7. PERIODIC MAINTENANCE AND CLEANING 7.1 PERIODIC MAINTENANCE The instrument does not require any periodical maintenance. 7.2 CLEANING  Clean the device when the chamber is at room temperature after disconnecting the power cable.  Clean the device with a damp cloth to remove dirt and dust. ...
  • Page 19: Error Codes Explanations

    9.1 ERROR CODES EXPLANATIONS The temperature sensor endings are broken. The error code flashes on the temperature display and an audible alarm sounds. An electronic failure occurs in the microprocessor. The error code flashes on the temperature display and an audible alarm sounds. The temperature sensor measures a temperature higher than 147°C.The error code is shown on the temperature display and an audible alarm sounds.
  • Page 20: Fuse Replacement

    9.2 FUSE REPLACEMENT The fuses shall always be replaced by the authorized personnel.
  • Page 21: Electrical Circuit Diagram

    ELECTRICAL CIRCUIT DIAGRAM 10.1 NB 5 / NB 9 / NB 20 ELECTRICAL CIRCUIT DIAGRAM...

This manual is also suitable for:

Nb 9Nb 20

Table of Contents