TEL: +(90) 312 399 28 30 (pbx) FAX: +(90) 312 399 21 97 E-mail: sales@nuve.com.tr WARRANTY CERTIFICATE Nüve warrants that the equipment delivered is free from defects during material and workmanship. This warranty is provided for a period of two years. The warranty period begins from the delivery date.
Page 3
Information contained in this document is the property of Nüve. It may not be duplicated or distributed without its permission. PLEASE REGISTER ONLINE TO VALIDATE WARRANTY: To register your warranty online, please visit our web page www.nuve.com.tr and fill in WARRANTY REGISTRATION FORM.
SECTION 1 INTRODUCTION 1.1 USE AND FUNCTION The EN 300, EN 400, EN 400 P, EN 500 and EN 500 P incubators are designed to incubate samples in biological, medical and pharmaceutical laboratories and in many industrial control laboratories. They maintain drying and incubation temperatures between 5°C above the ambient temperature and 80°C and keep the temperature stable within the given tolerances.
2.3 GENERAL PRESENTATION GLASS DOOR GLASS DOOR HANDLE VENTILATION HOLE DİSPLAY AND CONTROL PANEL SAFETY THERMOSTAT ADJUSTING BUTTON ON/OFF SWITCH CHAMBER GASKET SHELF DOOR SUPPLY INLET POWER CABLE...
SECTION 3 INSTALLATION PROCEDURE 3.1 LIFTING AND TRANSPORT Because of the heavy weight of the incubator, all lifting and transport must be carried out using proper handling equipment. The incubator must be supported from underneath and never turned over. 3.2 UNPACKING Remove the packing cardboard box and the second nylon packing around the incubator.
3.4 MAINS SUPPLY The incubator requires 230 V, 50 Hz. Please make sure that the supplied mains matches the required power ratings. Always plug-in the ıncubator to correctly grounded sockets. A supply fitted with a circuit breaker should be used for protection against indirect contact in case of an isolation fault.
4.2 DISPLAY AND CONTROL PANEL 01-Temperature Display This display shows the chamber temperature during the operation and the set temperature value during programming. The error codes are also shown on this display. 02-Time Display This display shows the elapsed time during the operation and the set value during programming.
4.3 PROGRAMMING SUMMARY Push the temperature setting button. Set the temperature by pushing the “value increase/decrease” keys on the temperature adjustment side. Push the “temperature set” key again to save the temperature value. Push the “time set” key. See “t ın” on the temperature display. Set the time value by pushing the “value increase/decrease”...
The incubator does not require any periodical maintenance which is carried out by the operator. Please contact to Nuve agent for an authorised service or maintenance 5.2 CLEANING After unplugging the equipment and the equipment is at the room temperature, wipe down the incubator chamber to remove any undesirable effects of the operation, for example spillage.
The chamber temperature exceeds 85°C. The temperature sensor (Fe-Const) endings are broken. IN CASE OF ANY ERROR, THE PROGRAM İS STOPPED AUTOMATİCALLY AND İMMEDİATELY. PLEASE CONTACT TO AN AUTHORIZED NUVE AGENT TO SEEK TECHNICAL HELP IF AN ERROR OCCURS.