Calibrating the Turb PLUS 2000, CAL Mode

The instrument was calibrated for the 0-100 NTU range and tested prior to leaving the factory.
Therefore, it is possible to use the instrument directly out of the box. Under normal conditions, re-
calibration is required once every three months. Quarterly calibration ensures performance within
accuracy specifications.
The EPA and ISO recommend that on-line turbidimeters be calibrated with a
primary standard at least once every three months if they are to be used for
reporting purposes.
Relay contacts and the 4-20mA will be frozen (held at the current state) while the instrument is in
the calibration mode. While in the calibration mode, the instrument has a time-out feature that
automatically returns the system operation to the AUTO mode after a fifteen (15) minute period of
inactivity.