Setup, Configuration and Calibration
6.2.5 Offset – Offset Reading Adjustment
A reading offset can be made to allow for the Turb PLUS 2000 to agree with another instrument if
desired. This range is limited to ± 1NTU. For more information see section 6.3.6.
To change the Offset:
1.
Enter the CONFIG mode by pressing the
illuminated, then press the
2. Press the
button to move to the next menu until OFST is displayed.
3. Select the offset using the
stay illuminated in AUTO as a reminder that it is being used.
4. Select the offset using the
5. Press the
button to accept the setting and move to the next menu or the
to AUTO.

6.2.6 Access Code - Setting Security

The Turb PLUS 2000 is equipped with a security access feature that can be activated. If this
option is turned ON, the user is required to input an access code into the instrument to get to any
mode other than AUTO. A three-digit code is used. When the CODE is turned On the icon will be
illuminated in AUTO as a reminder that an access code is being used.
Xylem Analytics Germany highly recommends the use of a security access
code in order to help prevent unauthorized access to the system and data.
To set the Security to on/off:
1. Enter the CONFIG mode by pressing the
illuminated, then press the
2. Press the
button to move to the next menu until CODE is displayed.
3. Use the
or
buttons to select the setting. (On or OFF).
100193 Rev 0 ba77254e01 11/2020
and
button.
or
buttons to turn the function On. When the OFST is turned ON it will
or
buttons.
and
button.
buttons until the arrow beside CONFIG is
buttons until the arrow beside CONFIG is
Turb PLUS 2000
button to exit and return