6.2.10 Display Units - Setting Displayed Units

The most common unit is NTU (Nephelometric Turbidity Units) however the instrument can
display in FNU (Formazine Nephelometric Units). All instruments are shipped from the factory set
in NTU mode.
To set/change the units:
1. Enter the CONFIG mode by pressing the
illuminated, then press the
2. Press the button to move to the next menu until UNIT is displayed.
3. Use the
or
buttons to select the setting NTU or FNU.
4. Press the
button to accept it and move to the next menu or the

6.2.11 Ultrasonic Cleaning - Cleaning Function

This allows for a selection menu to turn off the ultrasonic cleaning function if desired. The default
mode is On.
To set the ultrasonic cleaning to On/Off:
1. Enter the CONFIG mode by pressing the
illuminated, then press the
2. Press the
button to move to the next menu until CLN is displayed.
3. Select On or OFF using
4. Press the
button to accept I the setting and move to the next menu or the
to AUTO.
34
and
button.
and
button.
the
and
buttons.
Setup, Configuration and Calibration
buttons until the arrow beside CONFIG is
button to exit and return to AUTO.
buttons until the arrow beside CONFIG is
100193 Rev 0 ba77254e01
button to exit and return
11/2020