Xylem YSI Pro2030 Quick Start Manual

Xylem YSI Pro2030 Quick Start Manual

Advertisement

Quick Links

Pro2030
Quick Start Guide
This Quick Start Guide is meant to serve as a quick reference in calibrating and
operating the Pro2030. It is not intended to replace the information found in the Users
Manual.
Setting Dissolved Oxygen Sensor & Membrane Type
The instrument's dissolved oxygen (DO) sensor type must be configured for the sensor
installed. Failure to do this may result in damage not covered under warranty. The
correct DO membrane type must be configure for the membrane in use or the DO
readings will be erroneous.
The instrument will step you through an initial configuration when powered on for the
first time. This allows you to set the language, sensor and membrane options. Use the
up or down arrow keys to highlight the appropriate language, sensor, and membrane,
then press enter to confirm. Check the label on the box of the sensor and membrane
to determine which option you have. To change the sensor or membrane type after the
initial configuration, press the Menu key.
Installing the DO Sensor
Prior to use, the dissolved oxygen (DO) sensor must be installed on the Pro2030 cable.
Failure to install the DO sensor make cause permanent damage to the cable not
covered under warranty.
1.
Remove the red plastic plug from the cable's port by pulling it straight out of the
port. This can be discarded.
2.
Remove the red plastic plug from the DO sensor's connector by pulling it
straight off the sensor. This can be discarded.
3.
Ensure both the sensor connector and cable connector are clean and dry.
4.
Grasp the sensor with one hand and the cable bulkhead in the other.
5.
Push the sensor into the connector on the cable until it is firmly seated and only
1 o-ring is visible. Failure to properly seat the sensor may result in damage.
6.
Twist the sensor clockwise to engage the threads and finger tighten. Do NOT
use a tool. This connection is water-tight.
Item #605057

Advertisement

loading

Summary of Contents for Xylem YSI Pro2030

  • Page 1 Pro2030 Item #605057 Quick Start Guide This Quick Start Guide is meant to serve as a quick reference in calibrating and operating the Pro2030. It is not intended to replace the information found in the Users Manual. Setting Dissolved Oxygen Sensor & Membrane Type The instrument’s dissolved oxygen (DO) sensor type must be configured for the sensor installed.
  • Page 2 Installing the DO Membrane The DO sensor is shipped with a red protective cap covering the electrodes. A new cap membrane with electrolyte must be installed on the DO sensor prior to use. Prepare the electrolyte solution according to the instructions on the bottle. Remove and discard the red protective cap or used cap membrane from the sensor.
  • Page 3: Conductivity Calibration

    Turn the instrument on. If using a polarographic sensor, wait 10 minutes for the DO sensor to stabilize. Galvanic sensors do not require a warm up time. Ensure the barometer reading along the bottom of the display is accurate. Press and hold the Cal key for three seconds. If Quick DO Cal is enabled, highlight Dissolved Oxygen and press enter.
  • Page 4: Taking Measurements

    Press Enter to complete the calibration. Or, press Cal to cancel the calibration and return to the Run screen. ‘Calibration Successful’ will display for a few seconds to indicate a successful calibration and then the instrument will return to the Run screen. If the calibration is unsuccessful, an error message will display on the screen.