Setup, Configuration and Calibration
In case of an error in the Turb PLUS 2000, the 4-20 mA reading can be used to indicate a
problem by sending the current to either 4.00 mA, 2.00 mA. 0.00 mA, or OFF.
The factory default setting is OFF and the 4-20mA is unaffected by any error condition.
To change the Error Level:
1. Enter the CONFIG mode by pressing the
illuminated, then press the
2. Press the
button to move to the next menu until ERLV is displayed.
3. Select the desired ERLV by using the
and move to the next menu or the
6.2.2 Baud Rate - Selecting the RS–485 Baud Rate
The baud rate can be selected for the operation of the I/O port. Available baud rates are: 1200,
2400, 4800, 9600, or 19200.
To change the Baud Rate:
1. Enter the CONFIG mode by pressing the
illuminated, then press the
2. Press the
button to move to the next menu until BAUD is displayed.
3. Select the desired BAUD rate by pressing the
4. Press the
button to accept the setting and move to the next menu or the
to AUTO.
100193 Rev 0 ba77254e01 11/2020
and
button.
and
buttons then press the
button to exit and return to AUTO.
and
button.
buttons until the arrow beside CONFIG is
buttons until the arrow beside CONFIG is
or
buttons to change the displayed baud rate.
Turb PLUS 2000
button to accept the setting
button to exit and return
27