Abort Calibration - Xylem WTW Turb PLUS 2000 Operating Manual

Process turbidimeter
Table of Contents

Advertisement

Setup, Configuration and Calibration
4. To calibrate for 10 NTU and below only, press the
step 7.
5. Press the
button to initialize the calibration.
6. The lower display will count down from 30.
7. When complete, the lower display will now change to show alternating 10 and
NTU standard.
8. If the alternating 10 and
9. Insert the requested 10.0 NTU standard. Index the standard to the lowest value on the upper display.
10. Press the
button to initialize the calibration.
11. The lower display will count down from 60.
12. When complete, the lower display will now change to show alternating 02 and
NTU standard.
13. Insert the requested 0.02 NTU standard. Index the standard to the lowest value on the upper display.
14. Press the
button to initialize the calibration.
15. The lower display will count down from 30.
16. When complete the instrument will show CAL Good. Press the
briefly show STRT and then go to AUTO.

6.3.4 Abort Calibration

If at any time during the calibration process, you can abort the calibration by pushing theuntil
the screen displays ABRT then press the
The instrument will save and use any points that were calibrated successfully. It should be noted
100193 Rev 0 ba77254e01 11/2020
is not displayed, push theuntil this display is shown.
button. The instrument returns to AUTO operation.
button to bypass the 100 NTU and proceed to
button to accept. The display will
Turb PLUS 2000
, requesting the 10.0
, requesting the 0.02
45

Advertisement

Table of Contents
loading

This manual is also suitable for:

600026

Table of Contents