Pro1030
Quick Start Guide
This Quick Start Guide is meant to serve as a quick reference in calibrating and
operating the Pro1030. It is not intended to replace the information found in the User
Manual.
Installing the pH or ORP Sensor
Prior to use, install either a Pro Series pH, pH-amplified or ORP sensor. Failure to install
a pH or ORP sensor may cause permanent damage to the cable not covered under
warranty.
1.
Remove the temporary plastic plug from the cable's port by pulling it straight
out of the port. This can be discarded.
2.
Remove the plastic plug from the sensor's connector by pulling it straight off the
sensor. This can be discarded.
3.
Ensure both the sensor connector and cable connector are clean and dry.
4.
Grasp the sensor with one hand and the cable bulkhead in the other.
5.
Push the sensor into the connector on the cable until it is firmly seated and only
1 o-ring is visible. Failure to properly seat the sensor may result in damage.
6.
Twist the sensor clockwise to engage the threads and finger tighten. Do NOT
use a tool. This connection is water-tight.
7.
You may remove the shipping storage vessel from the sensing end of the
sensor. This can be saved for later use when storing the sensor for long periods
of time. For short term storage (<30 days), place the probe in the grey storage
sleeve with a sponge moistened with tap water.
pH Calibration
The Pro1030 pH sensor can be calibrated by performing a 1, 2 or 3-point calibration. At
least one of the calibration points must be done with pH buffer 7 or 6.86. For highest
accuracy, perform at least a 2-point calibration.
1 Point Calibration
1.
Place the sensor in pH buffer 7 or 6.86 and allow the temperature and pH
readings to stabilize.
2.
Press and hold Cal for three seconds.
3.
Highlight pH and press Enter. If pH is not listed as an option, check the System
Setup menu to ensure pH is enabled in the ISE Sensor Type menu.
Item #605183

Advertisement

loading

Summary of Contents for Xylem YSI Pro1030

  • Page 1 Pro1030 Item #605183 Quick Start Guide This Quick Start Guide is meant to serve as a quick reference in calibrating and operating the Pro1030. It is not intended to replace the information found in the User Manual. Installing the pH or ORP Sensor Prior to use, install either a Pro Series pH, pH-amplified or ORP sensor.
  • Page 2 Highlight 1 point and press Enter. If necessary, use the up and down arrow keys to adjust the pH buffer value. Press Enter to complete the calibration or press Cal to cancel. ‘Calibration Successful’ will display for a few seconds to indicate a successful calibration and then the instrument will return to the Run screen.
  • Page 3 ORP Calibration Place the clean sensor in ORP calibration solution. Wait for the ORP and temperature readings to stabilize. Press and hold Cal for three seconds. Highlight ORP and press Enter. If ORP is not listed as an option, check the System Setup menu to ensure ORP is enabled in the ISE Sensor Type menu.
  • Page 4 If the calibration is unsuccessful, an error message will display on the screen. Press the Cal key to exit the calibration error message and return to the Run screen. Refer to the troubleshooting section of the manual for a possible solution or contact YSI Technical Support.