Guidance and manufacturer's declaration - electromagnetic emissions
The CR-IR 355RU is intended for use in the electromagnetic environment specified below.
The customer or the user of the CR-IR 355RU should assure that it is used in such an environment.
Emissions test
Compliance
RF emissions
Group 1
CISPR 11
RF emissions
Class A
CISPR 11
Harmonic emissions
Class A
IEC61000-3-2
Voltage fluctuations/
flicker emissions
Complies
IEC61000-3-3
Guidance and manufacturer's declaration - electromagnetic immunity
The CR-IR 355RU is intended for use in the electromagnetic environment specified below.
The customer or the user of the CR-IR 355RU should assure that it is used in such an environment.
Immunity test
Electrostatic discharge
(ESD)
IEC61000-4-2
Electrical fast transient/
burst
IEC61000-4-4
Surge
IEC61000-4-5
Voltage dips, short
interruptions and
voltage variations of
power supply input
lines
IEC61000-4-11
Power frequency
(50/60Hz) magnetic
field
IEC61000-4-8
NOTE: U
is the a.c. mains voltage prior to application of the test level.
T
First Edtion 11.2005 897N0565
FCR XC-1 Operation Manual
The CR-IR 355RU uses RF energy only for its internal function.
Therefore, its RF emissions are very low and are not likely to cause any interference
in nearby electronic equipment.
The CR-IR 355RU is suitable for use in all establishments other than domestic and
those directly connected to the public low-voltage power supply network that
supplies buildings used for domestic purposes.
IEC 60601 test level
±6kV contact
±8kV air
±2kV for power supply
lines
±1kV for input/output lines
±1kV differential mode
±2kV common mode
<5% U
T
(>95% dip in U
)
T
for 0.5 cycle
40% U
T
(60% dip in U
)
T
for 5 cycles
70% U
T
(30% dip in U
)
T
for 25 cycles
<5% U
T
(>95% dip in U
)
T
for 5 s
3 A/m
Electromagnetic environment - guidance
Compliance level
±2kV contact
±4kV contact
±6kV contact
±2kV air
±4kV air
±8kV air
±2kV for power supply
lines
±1kV for input/output lines
±1kV differential mode
±2kV common mode
<5% U
T
(>95% dip in U
)
T
for 0.5 cycle
40% U
T
(60% dip in U
)
T
for 5 cycles
70% U
T
(30% dip in U
)
T
for 25 cycles
<5% U
T
(>95% dip in U
)
T
for 5 s
3 A/m
Electromagnetic environment -
guidance
Floors should be wood, concrete or
ceramic tile. If floors are covered with
synthetic material, the relative
humidity should be at least 30%.
Mains power quality should be that of
a typical commercial or hospital
environment.
Mains power quality should be that of
a typical commercial or hospital
environment.
Mains power quality should be that of
a typical commercial or hospital
environment. If the user of the
CR-IR 355RU requires continued
operation during power mains
interruptions, it is recommended that
the CR-IR 355RU be powered from
an uninterruptible power supply or a
battery.
Power frequency magnetic fields
should be at levels characteristic of a
typical location in a typical commercial
or hospital environment.