Portable and mobile RF communications equipment and the CR-IR 355RU
The CR-IR 355RU is intended for use in the electromagnetic environment in which radiated RF disturbances are
controlled.
The customer or the user of the CR-IR 355RU can help prevent electromagnetic interference by maintaining a minimum
distance between portable and mobile RF communications equipment (transmitters) and the CR-IR 355RU as
recommended below, according to the maximum output power of the communications equipment.
Rated maximum output power
of transmitter
W
0.01
0.1
1
10
100
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in metres (m)
can be estimated using the equation applicable to the frequency of the transmitter, where P is the maximum output
power rating of the transmitter in watts (W) according to the transmitter manufacturer.
NOTE 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations.
Electromagnetic propagation is affected by absorption and reflection from structures, objects and people.
First Edtion 11.2005 897N0565
FCR XC-1 Operation Manual
Recommended separation distances between
Separation distance according to frequency of transmitter
150 kHz to 80 MHz
d = 1.2 P
0.12
0.38
1.2
3.8
12
m
80 MHz to 800 MHz
800 MHz to 2.5 GHz
d = 1.2 P
0.12
0.38
1.2
3.8
12
1
d = 2.3 P
0.23
0.73
2.3
7.3
23