FujiFilm FCR XC-1 Operation Manual page 25

Table of Contents

Advertisement

The FCR XC-1 (CR-IR 355RU) can be network-connected to a Console or printer or the like.
Illustrated below is an example of system configurations for the connected for FCR XC-1 (CR-IR 355RU) and
Console.
Accessory equipment connected to the analog and digital interfaces must be in compliance with the respective
nationally harmonized IEC standards (i.e. IEC 60950 for data processing equipment, IEC 60065 for video
equipment, IEC 61010-1 for laboratory equipment, and IEC 60601-1 for medical equipment).
Furthermore all configurations shall comply with the system standard IEC 60601-1-1. Everybody who connects
additional equipment to the signal input part or signal output part configures a medical system, and is therefore,
responsible that the system complies with the requirements of the system standard IEC 60601-1-1. If in doubt,
consult the technical services department or your local representative.
2
■ System configuration example
FCR XC-1 (CR-IR 355RU)
CR Console (CR-IR 348CL)
DRYPIX 1000
HUB
2-3
First Edtion 11.2005 897N0565
FCR XC-1 Operation Manual

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Cr-ir 355ru

Table of Contents