Thermo Scientific 410iQ Instruction Manual page 78

Carbon dioxide analyzer
Table of Contents

Advertisement

Operation
Settings
3-52
410iQ Instruction Manual
Photodiode Current Too Low: Displays OK/Fail for Photodiode
Current Too Low.
Sensor Detected: Displays OK/Fail for sensor detected.
Sensor Communication: Displays OK/Fail for sensor
communication.
Valid Calibration: Displays OK/Fail for valid calibration.
Board Communication: Displays OK/Fail for board communication
status.
Power Supply: Displays OK/Fail of power supplies. Power supply
goes red if any voltages are outside their limits. No voltage rows
ever get highlighted.
2.5V Diagnostic: Displays current voltage readings. Alarm limits
are not changeable.
3.3V Diagnostic: Displays current voltage readings. Alarm limits
are not changeable.
5.0V Diagnostic: Displays current voltage readings. Alarm limits
are not changeable.
24V Diagnostic: Displays current voltage readings. Alarm limits
are not changeable.