Thermo Scientific
O
Sensor
The O
Sensor screen provides status and alarms related to the O
2
2
If an item being monitored goes outside the lower or higher alarm limit, an
alarm is activated.
Use the  and  buttons to move up and down and the  and 
buttons to move left and right.
Home Screen>Settings>Health Check>Status and Alarms>O
The O
Sensor screen contains the following information:
2
Across:
O
Value: Displays the current value for each item.
Low Alarm: Displays low alarm status for each item.
High Alarm: Displays high alarm status for each item.
Units: Displays units for each item.
Down:
Concentration: Displays the current O
reading. User can adjust low and high alarm limits.
Temperature: Displays the current temperature reading.
Temperature Open: Displays OK/Fail for temperature open.
Temperature Short: Displays OK/Fail for temperature short.
Sensor Element Function: Displays OK/Fail for sensor element
function.
Outside Operational Spec.: Displays OK/Fail for outside operational
specification.
Sensor: This column lists items associated with the O
2
Operation
Sensor.
2
Sensor
2
sensor.
2
sensor concentration
2
410iQ Instruction Manual