Troubleshooting
Troubleshooting Guide
6-2
410iQ Instruction Manual
Problem
Solenoid current out of
range (option)
Pump current out of range
No output signal (or very
low output)
Calibration drift
Non-linear response
Excessive response time
Alarm – Instrument
Temperature
Alarm – Bench
Temperature
Possible Cause
Action
Micro SD Card
If Micro SD card was just
Programming
replaced, re-install the old one. If
the problem is fixed, request a
replacement Micro SD card.
Sticking or damaged
Reset solenoid via
solenoid
Settings>Health Check>Status
and Alarms>Valve and Pump
Resets screen. If damaged,
replace solenoid valve block.
Damaged or dirty pump
Reset pump via Settings>Health
Check>Status and Alarms>Valve
and Pump Resets. Inspect and
refurbish pump. If pump motor is
damaged, replace pump.
No sample gas reaching
Check input sample flow.
the analyzer
Ruptured pump
Rebuild pump head.
diaphragm
Blocked sample capillary
Unplug power cord. Clean or
replace capillary.
Line voltage fluctuations
Check to see if line voltage is
within specifications.
Defective pump
Rebuild pump.
Clogged capillaries
Unplug power cord. Clean or
replace capillary.
Clogged sample air filter
Replace filter element.
Incorrect calibration
Verify accuracy of multipoint
source
calibration source gas.
Leak in sample probe line
Check for variable dilution.
Partially blocked sample
Unplug power cord. Clean or
capillary
replace capillary.
Hang up/blockage in
Change element.
sample filter
Low line voltage
Check to see if line voltage is
within specifications.
Fan failure
Replace fan if not operating
properly.
Dirty fan filter
Clean or replace filter.
Overheating PCB
Locate defective PCB reporting
the error and replace if needed.
Defective bench heater
Check 10K ohm thermistor and
replace heater if bad.