Thermo Scientific 410iQ Instruction Manual page 50

Carbon dioxide analyzer
Table of Contents

Advertisement

Operation
Calibration
3-24
410iQ Instruction Manual
Adjust Low Range Span Coefficient: User manually adjusts low range
span coefficient when in dual or auto range mode.
Adjusted Concentration: Read only. Shows adjusted concentration based
on adjusted span coefficient.
Adjusted High Range Concentration: Read only. Shows adjusted high
range concentration based on adjusted high range span coefficient
when in dual or auto range mode.
Adjusted Low Range Concentration: Read only. Shows adjusted low
range concentration based on adjusted low range span coefficient when
in dual or auto range mode.
Calibrate: Calibrates span coefficient by saving the newly adjusted span
coefficient.