Operation
Calibration
3-14
410iQ Instruction Manual
The Calibrate Span Coefficient menu contains the following information:
Edit Span Concentration: User enters the span concentration when in
single range mode.
Edit High Range Span Concentration: User enters the high range span
concentration when in dual or auto range mode.
Edit Low Range Span Concentration: User enters the low range span
concentration when in dual or auto range mode.
Current High Range Concentration: Read only. Current high range
concentration reading when in dual or auto range mode.
Current Low Range Concentration: Read only. Current low range
concentration reading when in dual or auto range mode.
Current Span Coefficient: Read only. Displays what the current user-set
span coefficient is when in single range mode.
Current High Range Span Coefficient: Displays what the current user-set
high range span coefficient is when in dual or auto range mode.
Current Low Range Span Coefficient: Displays what the current user-set
low range span coefficient is when in dual or auto range mode.
Calculated Span Coefficient: Read only. After the "Edit Span
Concentration" value is entered, the new calculated span coefficient is
displayed when in single range mode.
Calculated High Range Span Coefficient: Read only. After the "Edit
High Range Span Concentration" value is entered, the new calculated
high range span coefficient is displayed when in dual or auto range
mode.
Calculated Low Range Span Coefficient: Read only. After the "Edit Low
Range Span Concentration" value is entered, the new calculated low
range span coefficient is displayed when in dual or auto range mode.
Calibrate: When pressed, updates the coefficient and the concentration
should match the span concentration.