Operation
Calibration
3-14
410iQ Instruction Manual
The Calibrate Span Coefficient menu contains the following information:
Edit Span Concentration: User enters the span concentration when in
●
single range mode.
Edit High Range Span Concentration: User enters the high range span
●
concentration when in dual or auto range mode.
Edit Low Range Span Concentration: User enters the low range span
●
concentration when in dual or auto range mode.
Current High Range Concentration: Read only. Current high range
●
concentration reading when in dual or auto range mode.
Current Low Range Concentration: Read only. Current low range
●
concentration reading when in dual or auto range mode.
Current Span Coefficient: Read only. Displays what the current user-set
●
span coefficient is when in single range mode.
Current High Range Span Coefficient: Displays what the current user-set
●
high range span coefficient is when in dual or auto range mode.
Current Low Range Span Coefficient: Displays what the current user-set
●
low range span coefficient is when in dual or auto range mode.
Calculated Span Coefficient: Read only. After the "Edit Span
●
Concentration" value is entered, the new calculated span coefficient is
displayed when in single range mode.
Calculated High Range Span Coefficient: Read only. After the "Edit
●
High Range Span Concentration" value is entered, the new calculated
high range span coefficient is displayed when in dual or auto range
mode.
Calculated Low Range Span Coefficient: Read only. After the "Edit Low
●
Range Span Concentration" value is entered, the new calculated low
range span coefficient is displayed when in dual or auto range mode.
Calibrate: When pressed, updates the coefficient and the concentration
●
should match the span concentration.