Thermo Scientific 410iQ Instruction Manual page 109

Carbon dioxide analyzer
Table of Contents

Advertisement

Optical Bench Settings
Thermo Scientific
The Optical Bench Settings screen allows the user to manually or
automatically set the detector gain and initial S/R (sample/reference) value.
The S/R value should be calibrated prior to other instrument calibrations.
For more information, see Chapter 4, "Calibration".
Use the  and  buttons to move up and down and the  and 
buttons to move left and right.
Home Screen>Settings>Measurement Settings>Advanced Measurement
Settings>Optical Bench Settings
Home Screen>Settings>Measurement Settings>Advanced Measurement
Settings>Optical Bench Settings>Continue to Detector Gain
Calibration
The Detector Gain Calibration screen contains the following information:
Across:
Description: Defines the actions the user can do.
Detector Gain: User can manually set or automatically calibrate the
detector gain.
Operation
Settings
410iQ Instruction Manual
3-83