Problem
Alarm – Bench Temp
Thermistor Short
Alarm – IR Source Current
Low
Alarm – IR Source Current
High
Alarm – IR Detector Bias
Low
Alarm – IR Detector Bias
High
Possible Cause
Action
Defective Thermistor
Replace Thermistor as needed.
Defective DMC board
Replace board as needed.
Cable connection
Check the cable connection from
the IR Source to the DMC board.
Defective IR Source
Replace IR Source as needed.
Defective DMC board
Replace board as needed.
Cable connection
Check the cable connection from
the IR Source to the DMC board.
Defective IR Source
Replace IR Source as needed.
Defective DMC board
Replace board as needed.
Cable connection
Check the cable connection from
the IR Source to the DMC board.
Defective IR Source
Replace IR Source as needed.
Defective DMC board
Replace board as needed.
Cable connection
Check the cable connection from
the IR Source to the DMC board.
Defective IR Source
Replace IR Source as needed.
Defective DMC board
Replace board as needed.
Cable connection
Check the cable connection from
the IR Source to the DMC board.
Defective IR Source
Replace IR Source as needed.
Defective DMC board
Replace board as needed.
Troubleshooting
Troubleshooting Guide
410iQ Instruction Manual