Replacing The Cap Lamp - Thermo Scientific 3600 PDM Manual

Personal dust monitor
Table of Contents

Advertisement

Replacing the
Cap Lamp
To replace the cap lamp
Thermo Fisher Scientific
Figure 5–45. Mass transducer chamber
Figure 5–46. Mass transducer O-rings
3. Using the cotton swab, apply the silicone compound to the O-rings.
4. Install the PDM cover plate (refer to the instructions earlier in this
chapter for information on opening the case) and perform a leak check.
If the camp lamp assembly malfunctions, the entire cap lamp assembly
should be replaced. Tool/materials needed for this procedure:
Phillips screwdriver
Small flat-head screwdriver
7/64-inch hex wrench
WARNING Whenever the PDM case is opened for maintenance, you must
perform a case leak check after reassembling the case. ▲
1. Locate the top cyclone plate on the right-hand side of the PDM unit
assembly:
and use a 7/64-inch hex wrench to remove the four hex screws from the
top cyclone plate
(Figure 5–47
and
Figure
Maintenance and Calibration

Replacing the Cap Lamp

5–48).
PDM 3600 Instruction Manual
5-31