Starting a Second
Sample Run
To start a second manual
Second Sample
Thermo Fisher Scientific
Figure 2–24. Warming screen
Figure 2–25. First Sample Screen #1
While sampling, a second sample may be started, for example if the
conditions change or the employees moves to a different area, a second
manual sample run can be started running concurrently with the first
sample run.
1. In the First Sample Screen #1
sample run:
buttons simultaneously to display the Reset Sample 2 screen
26).
Figure 2–26. Reset Sample 2 screen
2. In the Reset Sample 2 screen, press the "A" button to begin a second
manual sample run. The second manual sample run will begin
immediately after pressing the "A" button. The PDM unit will not
perform another 30-minute warm-up period before beginning the
second manual sample run, and the Second Sample screen will display.
Note After 30 seconds of inactivity in the Reset Sample 2 screen, the PDM
will automatically display the First Sample Screen #1. ▲
The Second Sample Screen
cumulative exposure and elapsed time for the second manual sample run
Screen
(Figure
2–25), press the "A" and "B"
(Figure
2–27) displays the mass concentration,