Maintenance and Calibration

Periodic Maintenance

Periodic
Maintenance
Opening the
PDM Case
5-20
PDM 3600 Instruction Manual
Figure 5–30. Tilt Audit Pass/Fail status prompt
In addition to regularly scheduled maintenance, the unit may require other
maintenance, including:
Replacing the batteries
Replacing the cap lamp
Replacing the pump
Lubricating the mass transducer O-rings
Performing a flow calibration.
WARNING Do not operate the TEOM 3600 PDM if its case is damaged
or otherwise compromised. ▲
CAUTION TEOM 3600 PDM maintenance should only be performed by
trained, authorized personnel.
Some of these procedures require the PDM case to be opened. If the PDM
case is opened for any reason, a case leak check must be performed.
Some of the maintenance procedures require the user to remove the cover
of the case of the PDM unit. A Phillips screwdriver is required to remove
the cover of the unit.