3.2.7 Electrical data: single-phase pumps
Note:
We recommend using regional supply protection of the maximum ratings specified in
Table: Electrical data (single-phase pumps with Part Numbers ending in -903 or -906)
and
Table: Electrical data (single-phase pumps with Part Numbers ending in
Protection of a higher rating must not be used.
The dual-voltage, dual-frequency motor is designed for a single-phase electrical supply
and is suitable for 50 Hz or 60 Hz operation. The motor can be manually switched
between nominal supply voltages of 110‑120 V and 220‑240 V (refer to
configure the motor
start-up current shown in
ending in -903 or -906)
ending in -904)
lag fuse must be used to prevent unnecessary fuse failure during pump start-up (in
accordance with local and regional electrical codes). Within five minutes, as the oil in the
pump warms up, the current drawn will slowly reduce to the full load current specified
in
Table: Electrical data (single-phase pumps with Part Numbers ending in -903 or -906)
and
Table: Electrical data (single-phase pumps with Part Numbers ending in
Table 11 Electrical data (single-phase pumps with Part Numbers ending in -903 or -906)
Pump
RV3, RV5, RV8 and
RV12
Table 12 Electrical data (single-phase pumps with Part Numbers ending in -904)
Pump
RV3, RV5, RV8 and
RV12
Table 13 Recommended regional supply protection
Area
UK
Europe
USA
Japan
Electrical cables
Recommended cord sets and fuses for regional requirements.
A65201880_AA
A65201880_AA - Technical data
on page 27). When a cold pump is started, the motor will draw the
Table: Electrical data (single-phase pumps with Part Numbers
and
Table: Electrical data (single-phase pumps with Part Numbers
for up to several seconds. An appropriate rated and pre-approved time-
Nominal supply (V)
220-240
230-240
110
115-120
Nominal supply (V)
200
200-210
100
100-105
Frequency (Hz)
Power (W)
50
60
50
60
Frequency (Hz)
Power (W)
50
60
50
60
Voltage (V)
230
230
110
100
Page 22
-904).
Check and
-904).
Full load current (A)
450
3.4
550
3.0
450
6.8
550
6.9
Full load current (A)
450
4.2
550
4.1
450
8.3
550
8.0
Rating (A)
5
5
13
13
03/2021 - ©Edwards Limited