6.11 Test the motor condition

Test the earth (ground) continuity and the insulation resistance of the pump-motor, in
accordance with local regulations for periodic testing of electrical equipment.
The motor of single-phase RV pumps complies with IEC 1010V1. We recommend that, to
maintain compliance with IEC 1010‑1, the earth continuity is less than 0.1 W and the
insulation resistance is greater than 10 W.
If the motor fails these tests, the motor must be replaced.
A65201880_AA
A65201880_AA - Maintenance
Page 46
03/2021 - ©Edwards Limited