Procedure

Figure 3-3. Operator's Check Equipment Setup
1. Connect the equipment as shown in Figure 3-3, with the detector connected
to input A on the analyzer. Form a thru connection by connecting the
detector's RF input to the RF output of the source. Switch all of the
instruments' line powers ON and allow them to warm up.
2. If the 8757 series analyzer's system interface is not engaged (the analyzer's
status line displays
.
Y INTF ON
3. On the analyzer, press
and measuring input A.
4. Adjust the S OP and S AR frequencies on the source to include the
frequency range of the device under test (the bandpass lter).
Ag lent 85025A/B/D/E
), press
Y INTF OFF
SYSTEM
. he analyzer's channel 1 should be active
PRESET
5
MORE
WEEP MODE
5
Operat on
3-7