Mounting The Measuring Device; Required Tool; Installation - Endress+Hauser Liquiphant FTL64 Operating Instructions Manual

Vibronic point level switch for liquids in high-temperature applications
Hide thumbs Also See for Liquiphant FTL64:
Table of Contents

Advertisement

Installation

14
 11
Examples of support in the event of dynamic load
5.2

Mounting the measuring device

5.2.1

Required tool

• Open-ended wrench for sensor installation
• Screwdriver for electrical connection
5.2.2
Installation
Align the tuning fork with the marking
Use the marking to align the tuning fork in such a way that medium can run off easily and
deposit buildup is avoided.
Markings may include the following:
• Material specification, thread description or circle on the hexagonal nut or on the weld-
in adapter
• II symbol on the back of the flange
316L/G1
 12
Marking to align the tuning fork
Installing in pipes
• Flow velocities up to 5 m/s at a viscosity of 1 mm
Check for correct functioning in the event of other process medium conditions
• The flow will not be significantly impeded if the tuning fork is correctly aligned and the
marking is pointing in the direction of flow
• The marking can be identified while the device is installed
Liquiphant FTL64
2
/s(cSt) and density 1 g/cm
Endress+Hauser
A0042356
A0042348
3
(SGU)