Installation

10
 3
Handling the device
5
Installation
WARNING
L
Loss of protection rating if the device is opened in a wet environment.
Only open the device in a dry environment!
• Any orientation for device with short pipe up to approx. 500 mm (19.7 in)
• Vertical orientation for device with long pipe
• Minimum distance between the fork tip and the tank wall or pipe wall: 10 mm (0.39 in)
 4
Installation examples in a vessel, pipe or tank
1
Temperature spacer with gas-tight welded glass feedthrough for tank with insulation and/or high process
temperatures
5.1

Installation conditions

5.1.1

Pay attention to temperature

Difference in temperature between outer and inner side of flange must not exceed
60 °C (140 °F).