Led Module Vu120 (Optional) - Endress+Hauser Liquiphant FTL64 Operating Instructions Manual

Vibronic point level switch for liquids in high-temperature applications
Hide thumbs Also See for Liquiphant FTL64:
Table of Contents

Advertisement

Electrical connection
28
Terminal assignment
 28
2-wire NAMUR ≥  2.2 mA/≤  1.0 mA, electronic insert FEL68
Behavior of switch output and signaling
MAX
MIN
 29
Behavior of switch output and signaling, electronic insert FEL68
MAX DIP switch for setting MAX safety mode
MIN DIP switch for setting MIN safety mode
RD
LED red for alarm
YE
LED yellow, switch status
GN LED green, operational status, device on
The Bluetooth module must be ordered separately, including the required battery, for
use in conjunction with electronic insert FEL68 (2-wire NAMUR).
6.2.7

LED module VU120 (optional)

Supply voltage
U = 12 to 55 V
,
DC
U = 19 to 253 V
, 50 Hz/60 Hz
AC
IEC 60947-5-6
8,2 V DC NAMUR
+
MAX
>0,7
COM
>0,5
MIN
1
2
1
2
L- L+
RD
YE
GN
L+
L+
L+
L+
L+
Liquiphant FTL64
-
.
2.2...3.8 mA
2
1
L-
0.4...1.0 mA
L-
2
1
2.2...3.8 mA
2
1
L-
0.4...1.0 mA
2
1
L-
< 1.0 mA
2
1
L-
Endress+Hauser
A0036066
A0037694