5.4
5.4.1
The display and control unit can be removed. This enables the flexible setup
of the display and control unit at the workplace.
Tool:
Material:
Requirements
− The weighing pan and the associated components have not been set up.
− For a device with an analytical draft shield or flat glass draft shield:
Procedure
t
t
t
t
1
t

Removing the Display and Control Unit

Positioning the Display and Control Unit

1 Torx Allen key, T20
1 soft support base
The side panels and upper panel have not been fitted.
Turn the device on its side and place it on the soft support base.
To loosen the control unit's retainer: Use the Torx Allen key to remove
both screws.
Remove the control unit and re-insert both screws into
the threaded holes.
Pull the connection cable between the control unit and the weighing
module (1) out of the control unit's retainer to the required length.
Place the device back on the device base on a level surface.