Adjustments Required; Performance Testing Required - Keysight Technologies N9030B Installation Note

Pxa spectrum analyzer, option c35, 3.5 mm input connector retrofit kit
Hide thumbs Also See for N9030B:
Table of Contents

Advertisement

Adjustments, and Performance Verification Tests
Adjustments, and Performance Verification Tests
Calibration Software and specified test equipment is required to perform the adjustments and the
performance verification testing. Information on how to obtain this software can be found at
http://www.keysight.com/find/calibrationsoftware

Adjustments Required

Ad justments
None anticipated, but dependent on possible performance test results.

Performance Testing Required

The following performance verification tests are the minimum set required to ensure that this newly
installed option is functioning properly. Performing only these tests does not guarantee the
instrument meets all specifications.
Performance Verification Tests
Frequency Response (all available tests)
Residual Responses
A full calibration is required to assure the instrument meets all specifications
The end user must ultimately determine whether they want a full calibration to be performed after
the installation of this upgrade or not. If a full calibration is required, arrangements regarding the
level of calibration must be made between the end user and the calibration provider.
For assistance, contact your nearest Keysight Technologies Sales and Service Office. To find your
local Keysight office access the following URL, or if in the United States, call the following
telephone number:
http://www.keysight.com/find/assist
1-800-829-4444 (8 am - 8 pm ET, Monday - Friday)
8
Installation Note N9030-90078