Table 11. Guidance And Manufacturer's Declaration-Electromagnetic Immunity - Nonin Avant 9700 Operator's Manual

Digital pulse oximeter
Hide thumbs Also See for Avant 9700:
Table of Contents

Advertisement

Table 11: Guidance and Manufacturer's Declaration—Electromagnetic Immunity
Immunity Test
This device is intended for use in the electromagnetic environment specified below. The customer and/or
user of this device should ensure that it is used in such an environment.
Portable and mobile RF communications equipment should be used no closer to any part of the device,
including cables, than the recommended separation distance calculated from the equation applicable to
the frequency of the transmitter.
Conducted RF
3 Vrms
IEC 61000-4-6
150 kHz to 80 MHz
Radiated RF
3 V/m
IEC 61000-4-3
80 MHz to 2.5 GHz
a. Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land
mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with
accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey
should be considered. If the measured field strength in the location in which the device is used exceeds the
applicable RF compliance level above, the device should be observed to verify normal operation. If abnormal
performance is observed, additional measures may be necessary, such as reorienting or relocating the device.
b. Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.
NOTES:
• At 80 MHz and 800 MHz, the higher frequency range applies.
• These guidelines may not apply in all situations. Electromagnetic propagation is affected
by absorption and reflection from structures, objects, and people.
36
IEC 60601
Compliance
Test Level
Level
3 Vrms
3 V/m
Electromagnetic Environment—Guidance
Recommended Separation Distance
d
=
1.17 P
d
=
1.17 P
d
=
2.33 P
where P is the maximum output power rating of
the transmitter in watts (W) according to the
transmitter manufacturer and d is the
recommended separation distance in meters
(m).
Field strengths from fixed RF transmitters, as
determined by an electromagnetic site survey
should be less than the compliance level in
b
each frequency range.
Interference may occur in the vicinity of
equipment marked with the following symbol:
Technical Information