Display Unit Dip Switches - Nonin Avant 4100 Operator's Manual

Digital pulse oximetry system with bluetooth wireless technology
Table of Contents

Advertisement

Display Unit DIP switches

The Avant 4000 display unit contains eight DIP switches that are located behind the
back cover of the unit. The UP position is toward the top of the unit, and the
DOWN position is toward the bottom of the unit. The factory setting for all DIP switches
is the DOWN position.
Switch
Function
Switch 1
Reserved for Future Use
Switch 2
Alarm Disable Lock
Up—Alarm volume may be disabled
Down—Alarm volume cannot be disabled
Switch 3
Date Format
Up—International Date format (Day-Month-Year)
Down—U.S.A. Date format (Month-Day-Year)
Switch 4
Data Output
Up—Real-time (once per second) data output
Down—Print on demand
Switch 5
Normal / Fast Responding SpO
Up—Fast responding SpO
Down—Normal SpO
Switch 6
Normal / Slow SpO
Up—Slow Averaging (8 beat exponential average)
Down—Normal Averaging (4 beat exponential average)
Nurse Call Output
Switch 7
Up—Continuous
Down—Momentary
Switch 8
Factory / User-Defined Defaults
Up—User-Defined Defaults for Alarm Limits and Volume Settings
Down—Factory Defaults for Alarm Limits and Volume Settings
Output
2
serial output
2
serial output
2
and Pulse Rate Averaging
2
3
5
6
7
8
1
2
4
23

Advertisement

Table of Contents
loading

Table of Contents