Display is too dark or light
Lid will not close.
Error message appears.

11.1 Error Messages

Message
Cause
HEATER ERROR!
BLOCK ERROR!
ROOM ERROR!
ROOM > 30 °C
2. Faulty LCD panel
Backlight brightness is not
adjusted properly.
1. Foreign matter, between
lid heater
block.
2. Faulty
mechanism.
Refer
to
list
messages below.
Lid temperature sensor problem.
Lid temperature exceeds 132 °C.
Lid temperature can't reach the set
temperature.
Block temperature sensor or heating
element problems.
The
unit
can't
reach
temperature,
because
temperature is too high.
The block temperature exceeds 108
°C.
Room temperature sensor problem.
The room temperature exceeds 30
°C.
Two running units are too close to
each other.
32
2. Return
service.
Adjust Display
Potentiometer.
Remove the foreign object or
and sample
matter.
lid
lock
of
error
Check the nature of the error
and
take
actions.
Action
Reboot the unit.
Cool
temperature. Reboot the unit.
the
set
the
room
Reboot the unit
Cool
temperature. Reboot the unit.
Remove
blocking the vents.
Keep two running units at
least 30cm from each other.
the
unit
for
Brightness
the
suggested
down
the
room
down
the
room
any
objects