18 TECHNICAL SPECIFICATIONS
Anode current
(X-ray units with SW prior to 3.0):
3D
Pan / SmartPan
Ceph
Anode current
(X-ray units with SW 3.0 or later):
3D
Pan / SmartPan
Scanning ceph
Planmeca ProCeph
mAs range
mGy range
Linearity of radiation output
Cooling period
Exposure time:
3D
Pan
SmartPan
Scanning ceph
Planmeca ProCeph
78
Planmeca ProMax 2D & 3D s & 3D Classic
3D s / 3D Classic / 3D Mid: 1 - 14 mA ±10%
3D Max: 1 - 12.5 mA ±10%
1 - 16 mA ±10%
1 - 16 mA ±10%
3D s / 3D Classic / 3D Plus / 3D Mid:
1 - 14 mA ±10%
3D Max: 1 - 12.5 mA ±10%
1 - 16 mA ±10%
1 - 16 mA ±10%
16 mA ±10%
min. / max. as indicated ±(10% + 0.2 mAs)
min. / max. as indicated ±40%
< 0.1
Automatically controlled
3D s:
Pulsed, effective 4.8 - 36 s as indicated ±10%
3D Classic / 3D Plus / 3D Mid:
Pulsed, effective 2.4 - 36 s as indicated ±10%
3D Max:
Pulsed, effective 3.6 - 24 s as indicated ±10%
2D / 3D Classic:
2.7 - 16 s as indicated ±10%
2D tomography: 4 - 12 s as indicated ±10%
3D s / 3D Classic:
3.3 - 19 s as indicated ±10%
3D Plus / 3D Mid: 3.7 - 23 s as indicated ±10%
3D Max: 10 s as indicated ±10%
X-ray units with SW prior to 3.0:
Normal: 12 - 18.7 s as indicated ±10%
High Speed: 6.4 - 9.9 s as indicated ±10%
X-ray units with SW 3.0 or later:
6.4 - 9.9 s as indicated ±10%
0.1 - 0.8 s as indicated ±10%