9.2.3 Adjusting exposure values for current exposure
Child (XS)
Small adult (S)
Medium-sized adult (M)
Large adult (L)
Extra large adult (XL)
User's Manual (2D)
Factory presets for panoramic exposures
PATIENT SIZE
The exposure values have been preset at the factory for
each patient size. The preset exposure values are
average values and they are only meant to guide the user.
NOTE
FOR X-RAY UNITS WITH DIMAX SENSOR:
The preset exposure values are optimized for taking
exposures at enhanced resolution (Romexis setting). You
can use lower exposure values when taking exposures at
normal resolution.
NOTE
Always try to minimize the radiation dose to the patient.
The preset exposure values are shown in the following
table.
kV VALUE
62
64
66
68
70
If you need to adjust the preset exposure values for this
exposure:
1.
Select the kV / mA field.
2.
Use the minus or plus buttons to set the exposure
values you wish to use. To improve the image
contrast, reduce the kV value. To reduce the
radiation dose, reduce the mA value.
3.
Select the green check mark button.
4.
Select
a.
the forward button or
b.
the fast forward button if you want to skip the
next screen.
Planmeca ProMax 2D & 3D s & 3D Classic
9 PANORAMIC EXPOSURE
mA VALUE
5
6.3
8
10
12.5
29