Inspection and maintenance
3.6

Options

3.6.1

DAP measuring system

Measuring device for dose area product (Dose Area Product).
The DAP measuring device is used to measure the kerma area product (kerma = kinetic
energy released in matter) during an X-ray examination. It is primarily used to record the
radiation dose a patient has been exposed to.
The DAP is calibrated at the factory.
SIE
Function of the DAP measuring system
System is switched on.
Press the Test button on the DAP display.
X
Verify the function of the DAP measuring system by releasing an exposure.
3.6.2

Remote control

PMP
Battery replacement for remote control (yearly)
Replace the battery (9V alkaline).
SIE
Remote control function
Check the function of the remote control.
Switch on the light localizer of the collimator.
X
Initiate an exposure.
Siemens AG
Medical Solutions
The subsequently displayed value must be between 80-120 µGym
lution range 0.1 µGym) or 8-12µGym (DAP resolution range 0.01µGym) for a
high-resolution chamber.
If the display is incorrect, the measuring device must be calibrated. See the
"Replacement of parts" instructions (SPR8-230.841...).
A measured value is displayed in the DAP display as a function of the object in
the beam path.
SPR8-230.831.30.03.02