5 Installation
5.6

Connecting the L-509 to the Controller

NOTICE
Damage if an incorrect controller or motor cable is connected!
Connecting a stage to an unsuitable controller or using an unsuitable motor cable can cause
damage to the stage or controller.
 Only connect a stage to a suitable controller (p. 12).
 To connect the stage to the controller, only use a motor cable that is suitable for the
controller:
− C-663.11 controller: C-815.LSM1 motor cable
− C-663.12 controller: C-815.LSM2 motor cable
− SMC Hydra controller: C-815.LSH1 motor cable
− C-863 or C-884 controller: C-815.38 motor cable
Requirements
 You have read and understood the general notes on installation (p. 17).
 You have installed the controller.
 You have read and understood the user manual of the controller.
 The controller is switched off.
Tools and accessories
Cable from the scope of delivery of the stage (p. 11)
Suitable tools for tightening the screws to the connections

Connecting the L-509 to the controller

1. Remove the ESD protection from all connections of the L-509.
2. Connect the L-509, cable, and controller with each other.
3. Secure the connections with the integrated screws against accidental disconnection.
L-509 Stage
MP141E
Version: 1.1.0
31