Calibration Measurement; Checking The Accuracy Of The Blood Pressure Display - Omron HEM-9000AI Instruction Manual

Non-invasive blood pressure monitor with augmentation index
Hide thumbs Also See for HEM-9000AI:
Table of Contents

Advertisement

You will need:
(1) A mercury blood pressure monitor that is accurately adjusted (including a rubber bulb)*.
(2) T-tube
(3) 2 air tubes
(4) Something to wrap a cuff around such as a sturdy tube (strong material that is not
damaged or deformed by inflation)
* Even a mercury blood pressure monitor might cause an error in measurement due to a lack of mercury in the
mercury tank, dirt on the glass tube or clogging of the glass tube cap.
Connect the mercury blood pressure
1
monitor, the rubber bulb, the main unit
and the T-tube as shown in [Fig. 1],
to the right.
Press the F5 key [SETUP] to enter the
2
setup menu. [Fig. 2]
3
Move the
cursor to [METHOD] under
[ ]
[BP MEASUREMENT] in the "SETTING
LIST" column. [Fig. 3]

Calibration Measurement

Checking the Accuracy of the Blood Pressure Display

71
[Fig. 1]
[Fig. 2]
[Fig. 3]