Omron HEM-9000AI Instruction Manual page 79

Non-invasive blood pressure monitor with augmentation index
Hide thumbs Also See for HEM-9000AI:
Table of Contents

Advertisement

Error No
Error Name
Abnormal average
E82
value calculation
E85
Position Error
E86
Position Error
General Error Codes
Message Dialog
Main unit error.
Turn off main power supply switch.
Blood pressure monitor error.
Turn off main power supply switch.
AI pulse wave sensor unit error.
Turn off power supply and check AI pulse wave sensor unit.
AI pulse wave sensor unit error.
Prepare for measurement.
AI pulse wave sensor unit error.
Put pulse wave sensor in holder.
Replace clock battery.
Main Unit error. End of Calender. Unable to continue.
Turn off main power supply switch.
Printer error.
Turn off main power supply switch.
Printer error.
Add paper.
Printer error.
Paper jam. Remove jammed paper.
Printer cover is open.
Close cover.
Insert memory card.
There is no memory card inserted in the device.
Therefore the operation was denied.
Memory capacity is insufficient.
Replace with empty memory card.
Current memory card cannot be used.
This memory card cannot be used.
This system was not properly terminated.
The latest measurement data has not been saved.
When a memory card is ejected at this point, an unexpected
error may occur.Always eject memory card using the "SET UP"
screen menu.
This memory card has not been formatted. Push eject button,
remove memory card and insert a new memory card.
ID DATA has not been copied.
This value is out of range.
Measurement error.
The measurement data has not been saved.
AI pulse wave measurement unit error.
Turn off power supply and connect AI pulse wave measurement unit.
Error Message
AI pulse wave sensor unit
error.
Cannot calculate average
AI.
AI pulse wave sensor unit
error. Slide the AI pulse
wave sensor unit in the
indicated direction and start
measurement.
AI pulse wave sensor unit
error. Slide the AI pulse
wave sensor unit in the
indicated direction and start
measurement.
Error Content
An irregular pulse wave or
similar problem prevented
the average value from
being measured.
An artery is not under the
sensor.
An artery is not under the
sensor.
Action
Turn off the power.Turn the power back on. If the problem is not
resolved, contact the Omron Healthcare Professional Services
Turn off the power.Turn the power back on. If the problem is not
resolved, contact the Omron Healthcare Professional Services
Turn off the power. Check the pulse wave sensor unit connection
and/or replace the AI pulse wave sensor unit.
Measurement began while the AI pulse wave sensor was stored.
Follow the instructions for measurement preparation on pages 24-26.
The AI pulse wave sensor has been out of storage for an extensive period of time.
While holding down the slide stopper, place the AI pulse wave sensor in its original position.
Contact the Omron Healthcare Professional Services
The calendar function cannot be used after December 31, 2030.
Turn off the power. Turn the power back on. If the problem is not
resolved, contact the Omron Healthcare Professional Services
Open the printer cover and load paper. Refer to page 21.
Open the printer cover and remove the paper. Refer to page 21.
Close the printer cover.
Memory card has not been inserted. Insert a memory card into
the slot on the right side of the main unit.
Memory card has not been inserted. Insert a memory card into
the slot on the right side of the main unit.
Eject the memory card and replace it with a new one. (Page 67)
Turn off the power. Reinsert the memory card and turn the power
back on.
Format the memory card (page 68) or replace it with a different
memory card.
Data was not saved since the power was disconnected or
interrupted during measurement.
Turn off the power. Reinsert the memory card and turn the power
back on.
Obtain a new memory card and insert it into the memory card slot
on the main unit.
Try to copy the information again following the procedure on page
67 starting from the step on copying ID information.
The number just entered was not appropriate. Reenter the
number.
Retake the measurement.
Turn off the power. After double-checking the cable connection,
turn the power back on.
79
Error List
Action
• There is a possibility that the AI pulse wave
sensor is not on the artery. After reviewing
"How to Apply the AI Pulse Wave Sensor
Unit" (pages 24-26) and "AI Measurement"
(pages 38-41), retake the measurement.
• Slide the AI pulse wave sensor unit
one or two scales in the indicated
direction and restart the
measurement. For complete
information refer to page 30.
• Slide the AI pulse wave sensor unit
one or two scales in the indicated
direction and restart the
measurement. For complete
information refer to page 30.