Lauda P 10 Operation Manual page 109

Pro bath thermostats and circulation thermostats with base remote control
Hide thumbs Also See for P 10:
Table of Contents

Advertisement

Code
English Edition
31
SWV COMMAND 1 OLD
32
SWV RESERVED
33
MISSING ADAPT2
34
PRESS SWITCH2
35
KLIXON2
36
SMIN TOO SMALL
37
VALVE NOT CLOSED
38
CALIBRATION
39
VALVE CHANGED
40
WRONG NET FREQUENCY
41
WRONG NET VOLTAGE
42
NO PRO TYPE
43
NO PRO VOLTAGE
44
KM CURRENT NOT OK
45
STACK OVERFLOW CAN
46
STACK OVERFLOW ADC
47
STACK OVERFLOW CHECK
48
STACK OVERFLOW COOL
49
STACK OVERFLOW FAN
50
RTOS MISC
51
KM OFF AT SERV
52
Free 52
53
Free 53
54
Free 54
55
Free 55
56
Free 56
V04REV14
PRO bath thermostats and circulation thermostats
Description
Operating system 1 (command or base) software version
is too old
Reserved for software version XXX is too old
Pressure switch KM2 has triggered
KM2: to1 is too high, Klixon has most probably triggered
to1 is too low in minimum position
Cooling valve does not close correctly during adaptation
cycle
Faulty calibration of a temperature probe
A valve was modified/replaced
Invalid PRO device type
Invalid PRO mains voltage setting
Refrigerating machine frequency incorrect
TaskCan: 90 % stack utilization exceeded
TaskADC: 90 % stack utilization exceeded
TaskCheck: 90 % stack utilization exceeded
TaskCool: 90 % stack utilization exceeded
TaskFan: 90 % stack utilization exceeded
Refrigerating machine switched off during adaptation/
flushing
109 / 130