Fig. 31: Base menu, Part 2
5.8

Setting overtemperature protection Tmax

Fig. 32: Setting Tmax
V04REV14
Functions that cannot be executed are omitted from the menu structure.
WARNING!
Risk of heat transfer liquid overheating due to incorrect entry of
overtemperature switch-off point Tmax
Fire
Set the overtemperature switch-off point to 5 K above the
upper limit of the temperature range for your application.
The overtemperature switch-off point must be below the
flash point of the heat transfer liquid.
The maximum temperature Tmax for the overtemperature protection is set
at the rotary knob using a screwdriver. The exact preset maximum tem-
perature also appears automatically on the display.
Ä Chapter 7.7 "Checking the overtemperature protection"
on page 101