7.4
Temperature setpoint setting
The setpoint is the temperature, which the thermostat should reach and maintain constant.
or
Wait 4 seconds or
7.5
Automatic Shutdown Timer
The automatic Shutdown Timer can be activated/deactivated, read out or be set. At shutdown time, the pump, heater
and compressor are deactivated, the display then shows "StbY" ( 7.10).
1x
or
09/2020 j
SEt
Press key to show
Press, display flashes.
Decrement or increment the Temperature Setting ( 7.3).
(Temperature can be set from -25 °C up to 100 °C depending on
limits "Hi" and "Lo").
Display flashes 4 seconds new value is automatically accepted, or
value is accepted immediately with <Enter> key.
When the setpoint temperature is set below the bath temperature, it may take
up to one minute before the blue LED
Entering value closer than 5 °C to the maximum or minimum temperature limit
(Hi and Lo) are not accepted. A buzzer signal appears. If necessary, first
change the temperature limits ( 7.6).
ASd
Call the Automatic Shut-Down Timer
to
ASd
Read Timer:
The display shows... "00.00":
The display shows... "hh.mm" (any value):
hh.mm
(
minutes remaining to shutdown)
Set Timer:
Increment or decrement "hh.mm" (maximum 99:59)
Within 4 seconds after the last change of values confirm the timer value by press-
ing <Enter>.
You then proceed to Main Menu "Bath Temperature": A blinking point indicates
that the timer is running!!
No change is done without confirmation by <Enter>!
You can deactivate the timer any time by setting the value "00.00".
SEt
(Setpoint).
ASd
Alpha
lights.
.
timer not active;
timer active,
29 / 54