Heat transfer liquid, water
60 / 130
Table 8: Heat transfer liquid order numbers
Designation
Kryo 95
Kryo 60
Kryo 51
Kryo 40
Kryo 30
Kryo 20
Aqua 90
Ultra 350
Ultra 240
Therm 250
Therm 180
Therm 160
When using Kryo 30:
The water content decreases during longer periods of operation at
higher temperatures and the mixture becomes combustible (flash point
119°C). Check the mixing ratio using a hydrometer.
When using Aqua 90:
Evaporation losses occur at higher temperatures. In this case, use a bath
cover.
When using Ultra 350:
Do not use in combination with an EPDM hose.
Never use silicone oil in silicone tubes.
When using mineral oils:
Do not use in combination with an EPDM hose.
The proportion of alkaline earth ions in the water must be between
0.71 mmol/L and 1.42 mmol/L (equivalent of 4.0 and 8.0 °dH). Harder
water leaves limescale deposits in the device.
The pH value of the water must be between 6.0 and 8.5.
Distilled, deionized and demineralized water are unsuitable due to their
reactivity. High-purity water and distillates are suitable as a heat transfer
liquid after 0.1 g of soda (Na
liter of water.
Sea water is unsuitable due to its corrosive properties.
Avoid chlorine content in the water at all costs. Do not add chlorine to
the water. Cleaning agents and disinfectants, for example, contain
chlorine.
PRO bath thermostats and circulation thermostats
Container size
Cat. No.
5 L
10 L
LZB 130
LZB 230
LZB 102
LZB 202
LZB 121
LZB 221
LZB 119
LZB 219
LZB 109
LZB 209
LZB 116
LZB 216
LZB 120
LZB 220
LZB 107
LZB 207
LZB 108
LZB 208
LZB 122
LZB 222
LZB 114
LZB 214
LZB 106
LZB 206
CO
, sodium carbonate) is added for every
2
3
20 L
LZB 330
LZB 302
LZB 321
LZB 319
LZB 309
LZB 316
LZB 320
LZB 307
LZB 308
LZB 322
LZB 314
LZB 306
V04REV14

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents