Section 6-3
Monitor Adjustment
The Monitor can be tilted or rotated.
Tilt angle: up/down 11°
Rotation angle: right/left 90°
The digital control panel is located at the front of the color monitor. If you are not satisfied with the factory
settings, use these controls to program those you prefer in each resolution. Then, these adjusted
settings are kept in memory even if you change the resolution or turn off the monitor.
Decrease
(Left)
Menu key:
This button will enable the On Screen Display (OSD-instructions). This button is also used to select
the function in the Main Menu or to save the settings in the Sub Menu. (Push for approx. 4 sec.)
Decrease (Left):
Use this button to move down the OSD selection menu and adjust the attribute of the monitor while
in OSD menu. Pressing this button out of the OSD menu allows you to decrease the level of
contrast of the display screen.
Increase (Right):
Use this button to move down the OSD selection menu and adjust the attribute of the monitor while
in OSD menu. Pressing this button out of the OSD menu allows you to increase the level of contrast
of the display screen.
6-3-1
Monitor Calibration
The procedure consists of 3 parts:
1.) Degauss
2.) Color Temp - Calibration
3.) Contrast and Brightness
4.) Geometry
5.) Convergence
6.) Color temperature
7.) B-Mode Quality
6-2
3
DRAFT (A
Increase
Menu
(Right)
Figure 6-1 Monitor Adjustment buttons
Section 6-3 - Monitor Adjustment