GE Voluson 730Expert Service Manual page 113

Table of Contents

Advertisement

GE H
- K
EALTHCARE
RETZTECHNIK
D
105899, R
IRECTION
EVISION
4-4-3
M Mode Checks
Table 4-4
Step
1
2
3
4
5
6
7
8
9
10
11
For further details refer to the Voluson® 730Expert Basic User Manual, Chapter 6, M Mode.
3
DRAFT (A
Figure 4-9 M Main and M Sub Menu
M Mode Functions
Task
Cursor Position
Adjust the M Cursor position with the TRACKBALL in the 2D Single image.
Activation of M Mode
Press the right or left trackball key to activate both Modes (2D/M).
M Mode Gain
Rotate the M MODE key to adjust the sensitivity (brightness) of the entire M image.
M Mode Depth
Common with 2D Mode Depth.
SPEED
By touching up or down, four different sweep speeds can be selected.
INVERT
Invert of the M Mode image. (Function is only available with endovaginal probes.)
FREQUENCY
Common with 2D Mode Frequency.
FORMAT
For selection of three different rations of display format.
DYN.CONTR.
Dynamic Range enhances a part of the grayscale to make it easier to display pathology.
ENHANCE
Due to this function a finer, sharper impression of the image is produced.
REJECT
It determines the amplitude-level below which echoes are suppressed (rejected).
Chapter 4 - Functional Checks
29, 2008)
PRIL
Expected Results
V
® 730E
OLUSON
XPERT
S
M
ERVICE
(BT03)
ANUAL
4-11