GE Voluson 730Expert Service Manual page 322

Table of Contents

Advertisement

GE H
- K
EALTHCARE
RETZTECHNIK
D
105899, R
IRECTION
EVISION
10-7-9-3
No Meter Probe Adapter Procedure
Follow these steps to test each transducer for leakage current.
1.) Turn the Voluson® 730Expert unit OFF.
2.) Plug the unit into the test meter, and the meter into the tested AC wall outlet.
3.) Plug the external probe into the meter's "EXTERNAL" connector.
4.) Set the meter's "FUNCTION" switch to EXTERNAL position.
5.) Connect the probe for test with the connector of the console.
6.) Add the saline probe and the imaging area of the probe into the saline bath.
Make sure not to immerse the probe into the liquid beyond the immersion level.
CAUTION
!! CAUTION:
7.) Have unit power ON for the first part; turn it OFF for the second half.
8.) Depress the ISO TEST rocker switch and record the highest current reading.
9.) Follow the test conditions described in
10.)Keep a record of the results with other hard copies of Preventive Maintenance (PM) data or
corrective actions taken.
10-22
3
DRAFT (A
Figure 10-6 Check Without Probe Adapter
Section 10-7 - Electrical Safety Tests
29, 2008)
PRIL
Table 10-19
for every transducer.
V
® 730E
OLUSON
XPERT
S
M
ERVICE
(BT03)
ANUAL