GE Voluson 730Expert Service Manual page 259

Table of Contents

Advertisement

GE H
- K
EALTHCARE
RETZTECHNIK
D
105899, R
IRECTION
EVISION
8-9-2-5
Restore all Exams (of the "old" system) to the "new" system (cont'd)
The folder which is selected to be renamed (e.g., A09008, that contains the backup data from a different
NOTICE
system - in this case the backup data of the
!! NOTICE:
Thus, the same data can be shared between two systems by renaming the respective backup folders
to the serial number of the accessing system (in this case the
5.) Close the "Settings" window with the OK button.
6.) Click on the OPEN icon on the upper left side of the screen and choose "Drive Network".
7.) Using the
8.) To select all available exams, click on the SELECT TO END button.
9.) Click on the BACKUP / RESTORE button.
Open
button
"Drive Network"
all exams are selected
Backup / Restore
10.)In the "Backup / Restore" window click on the RESTORE button.
If an exam is about to be restored that already exists on the hard disk of the
NOTICE
shows the Patient Name and Patient ID and asks for the action to be taken.
!! NOTICE:
3
DRAFT (A
, and the right trackball key SET, select the first exam of the list.
TRACKBALL
button
Figure 8-26 Step 5.) to 9.)
Figure 8-27 Identical exam already exists
Chapter 8 - Replacement Procedures
29, 2008)
PRIL
"old"
system), it is not copied, but simply renamed.
"new"
V
® 730E
OLUSON
XPERT
S
ERVICE
system).
"new"
system, a dialog
(BT03)
M
ANUAL
Select
to End
button
Close
button
8-19

Hide quick links:

Advertisement

Table of Contents
loading