After each installation, the leakage currents have to be measured according to
WARNING
WARNING
IEC 60601-1 respectively UL 60601-1.
!! WARNING:
Please observe that some printers may not be medical devices! If the Bluetooth Printer and/or
CAUTION
Line Printers are not medical devices, they have to be located outside of the patient
!! CAUTION:
environment (according to IEC 60601-1 / UL 60601-1).
3
DRAFT (A
(cont'd)
below outlines Voluson® 730Expert hardware installation procedures described in the sub-
Connection Procedures
Monitor Connection
Foot Switch Connection
ECG-preamplifier Connection
Global Modem Connection
S-VHS Video Recorder Connection
B/W Video Printer Connection
Line Printer Connection
Digital Color Printer Connection
Bluetooth Printer Connection
External USB-Devices
1,5 m
1,5 m
Chapter 3 - Setup Instructions
29, 2008)
PRIL
Description
2,5 m
V
® 730E
OLUSON
XPERT
S
M
ERVICE
Page Number
3-10
3-10
3-11
3-12
3-13
3-15
3-16
3-17
3-18
3-20
1,5 m
(BT03)
ANUAL
3-9