Function Description Related To Indicated Values; Indicated Values; Digital Zero - Unipulse TM201 Operation Manual

Torque monitor
Hide thumbs Also See for TM201:
Table of Contents

Advertisement

5 Function Description Related to Indicated Values
5
Function Description
Related to Indicated Values
5-1.

Indicated values

Torque
Analog signals (±5V) from UTMⅡ are displayed as torque.
Rotation speed
Pulse signals from UTMⅡ are displayed as rotation speed.
Power
Power calculated from torque and rotation speed are displayed.
Chapter
5
5-2.

Digital zero

This is a function to set the torque readings to zero.
The digital zero reset function that releases digital zero is also included.
■ Execution procedure
This button cannot be clicked in state of standby for the trigger condition or during waveform
acquisition.
38
38
Power [W] = 2π X torque [Nm] X rotation speed [rpm] / 60
Key points
Based on these data, torque/power attributes of the rotation speed can be
obtained.
1.
Switch the PC application screen to the
"Waveform" tab.
2.
Click the "Digital Zero/Digital Zero Reset"
button.
Key points
When digital zero reset is performed, the condition before digital zero correction
is restored.

Advertisement

Table of Contents
loading

Table of Contents