Procedures Of Equivalent Input Calibration And Actual Load Calibration (Torque) - Unipulse TM201 Operation Manual

Torque monitor
Hide thumbs Also See for TM201:
Table of Contents

Advertisement

4-3.
Procedures of equivalent input calibration
and actual load calibration (torque)
Equivalent input calibration and actual load calibration are performed in the following procedures,
and check "Detail" to display detail items.
(1)
Calibration setting protection
*
(2)
Unit (torque)
(3)
Min. scale division (torque)
*
(4)
Calibration setting PC → TM201
*
(5)
Zero calibration
*
(6) a.
Equivalent
input calibration
(7)
Calibration setting protection
*
... Be sure to perform the setting.
(1) Release of calibration setting protection
Calibration setting protection is set to OFF.
Only this setting is performed in the "Management" tab.
<Setting value>
29
(6) b.
Actual load
calibration
or
OFF
Setting values are rewritable.
ON
Setting values are not rewritable.
Calibration setting protection that prevents calibration
is turned OFF ("Management" tab).
The unit is set according to UTMⅡ used.
Minimum value of the digital change desired is set.
(Optional if there is no change.)
Calibration setting is written to the TM201.
UTM Ⅱ is set to unloaded condition (where no torque
load is being applied to UTM Ⅱ) and the zero point is
registered.
Equivalent input calibration
Rated output value and rated capacity value of UTMⅡ
are registered.
Actual load calibration
The known torque load is applied to UTM Ⅱ and the
torque is registered.
Calibration setting protection is turned ON to avoid
operational errors ("Management" tab).
4 Calibration
Chapter
4
29